14

Performing a Coating Zero

1) Press the MULTI MODE key once to activate the measurement mode

options.

2) Use the UP and DOWN arrow keys to scroll through the sub menu items

until Coating Only (CT) mode is highlighted.

3) Press the ENTER key to select the measurement mode and return to the

measurement screen.

4) Apply a drop of couplant on the transducer and place the transducer in

steady contact with the probe zero disk (battery cover) and obtain a steady

reading.

Note: The coating measurement displayed will potentially be a value greater or

less than 0.

5) Press the MENU key once to activate the menu items tab. Press the MENU

key multiple times to tab right and the ESC key multiple times to tab left until

the PRB menu is highlighted and displaying the submenu items.