Dakota Ultrasonics

120

Setting the Damping Value

1) Press the MENU key once to activate the menu items tab. Press the MENU

key multiple times to tab right, and the ESC key multiple times to tab left,

until the TUNE menu is highlighted and displaying the submenu items.

2) Use the UP and DOWN arrow keys to scroll through the sub menu items

until DAMPING is highlighted.

3) Use the LEFT and RIGHT arrow keys to scroll through the DAMPING values

until the correct value is displayed to the right of the DAMPING menu item.

4) Finally, press the MEAS key to return to the measurement screen and begin

taking readings.

10.10 Auto Find

The DFX-7 is equipped with a very handy feature for locating the detection point of a
back wall echo that is not currently in the viewable region of the display. Suppose
the DFX-7 is currently setup with a delay of 0.0” and a width of 1.0”, and the user is
measuring material thickness of 2.5”. The DFX-7 will make the correct measurement
and display it digitally. However, the waveform will not show up at all in both A-Scan
views, and incorrectly in the B-Scan view without adjusting the range of the Delay
and Width accordingly.