beautypg.com

Traceable dissolved oxygen pen instructions – Control Company 4319 DISSOLVED OXYGEN PEN User Manual

Page 2

background image

TRACEABLE

DISSOLVED

OXYGEN

PEN

INSTRUCTIONS

®

AUTOMATIC SHUTOFF (Disable)
If no button has been pressed for ten (10) minutes,
the Automatic Shutoff feature will turn the meter off
to preserve battery life.

The automatic shutoff feature setting is overridden
while in the Record Mode. (See the “Record
Mode” section)

To disable the Automatic Shutoff feature, place the
unit into the Record Mode (see the “Record Mode”
section).

PROBE MAINTENANCE
Electrolyte Solution Replacement
If the calibration check does not return the desired
results (see the “Calibration Check” section) or if
the readings are unstable, it may be necessary to
replace the electrolyte solution in the probe head.

1. Unscrew the probe head.

2. Pour out the old electrolyte solution from the
probe head.

3. Fill the probe head with fresh electrolyte
solution.

4. Replace the probe head and tighten securely,
do not overtighten.

Diaphragm Replacement
A damaged diaphragm can cause the calibration
check to return incorrect results (see the “Calibration
Check” section) or cause the readings to be
unstable or inaccurate. The diaphragm may be
damaged if it comes into contact with a solid
object, or is subjected to impacts. The diaphragm
is integrated into the probe head.

1. Unscrew the probe head.

2. Pour out the old electrolyte solution from the
probe head.

3. Fill the replacement probe head with fresh
electrolyte solution.

4. Place the new probe head on to the meter and
tighten securely, do not overtighten.

ALL OPERATION DIFFICULTIES
If this meter does not function properly for any
reason, replace the batteries with a new, high
quality batteries (see the “Battery Replacement”
section). Low battery power can occasionally
cause an number of “apparent” operational
dif culties. Replacing the batteries with new fresh
batteries will solve most dif culties.

BATTERY REPLACEMENT
An erratic display, faint display,
no display, or a battery symbol
appearing on the display are all
indicators that the batteries need
replacement. The battery cover
is located at the bottom of the
unit. While holding the unit upside
down, unscrew the battery cover
by turning it counter-clockwise.
Remove the exhausted batteries
and replace them with four (4)
new AAA alkaline batteries. Make
certain to insert the new batteries
with the proper polarity as indicated by the “+” and
“–” symbols in the battery compartment. Replace
the battery cover and tighten securely.

ACCESSORIES

Cat. No. 7790– Replacement Probe Head
w/Diaphragm

Cat. No. 8024– Electrolyte Solution

WARRANTY, SERVICE, OR RECALIBRATION
For warranty, service, or recalibration, contact:

CONTROL COMPANY

4455 Rex Road

Friendswood, Texas 77546 USA

Ph. 281 482-1714 • Fax 281 482-9448

E-mail [email protected] • www.control3.com

Control Company is ISO 9001 Quality-Certi ed by DNV and

ISO 17025 accredited as a Calibration Laboratory by A2LA.

Control Company Cat. No. 4319

Traceable

®

is a registered trademark of

Control Company.

©

2009

Control Company.

92-4319-00 Rev. 0 111309