beautypg.com

Control Company 3250 MULTIMETER User Manual

Control company, Traceable, Multimeter instructions

background image

Traceable

®

is a registered trademark of Control Company ©2009 Control Company. 92-3250-00 Rev. 1 031709

Specifications
AC Voltage

Range

Resolution

Accuracy

3V

1mV

±2.0% + 3 digits

30V

10mV

±2.0% + 3 digits

300V

100mV

±2.0% + 3 digits

750v

1V

±2.0% + 4 digits

DC Voltage

Range

Resolution

Accuracy

300mV

100uV

±1.2% + 2 digits

3V

10mV

±0.5% + 2 digits

300V

100mV

±1.2% + 2 digits

750v

1V

±1.5% + 3 digits

Resistance

Range

Resolution

Accuracy

300Ω

0.1Ω

±1.2% + 3 digits

3KΩ

1Ω

±1.2% + 3 digits

30KΩ

10Ω

±1.2% + 3 digits

300KΩ

100Ω

±1.2% + 3 digits

3MΩ

1KΩ

±1.2% + 3 digits

30MΩ

10KΩ

±3.0% + 5 digits

Continuity

Range

Frequency

Accuracy

Buzzer

4.1KHz

<20Ω ±10Ω

Diode

Max

Resolution

Accuracy

0.8mA

1mV

±8.0% + 2 digits

Battery

Range

Resolution

Load

1.5V

1mV

1.5mA

9V

10mV

9mA

Maximum display:

1999

Range:

Auto-range

Indicator:

Bar graph

Maximum Display Count: 3200
Display:

LCD, 4-digit ⅝”

Features:

Data hold, low battery alert, auto power off

Power:

Two 1.5V batteries

Size:

4¼ x 2½ x ⅜ inches

Accessories red and black test leads, red and black alligator clip leads,
Supplied:

batteries, protective case, Traceable

®

Certificate, instructions

Figure 1

Description

1.

Display

2.

DATA HOLD/RESET Button

3.

Function Dial

4.

Positive Jack

5.

Negative Jack

6.

Continuity Buzzer Sound On/Off

7.

AC Voltage

8.

DC Voltage

9.

Battery test

10. Continuity/Diode
11. Resistance

PROBES
Select either the probe test leads or the clip leads. Plug the red test leads into the
positive “+” jack and the black test lead into the negative “-” jack (4 and 5 Fig. 1).

AUTO-OFF
If the reading has not changed for ten minutes the unit provides warning beeps and
then automatically turns off to conserve battery life. To turn the unit back on and
resume readings press the RESET button (2, Fig. 1).

ACV/DCV/BATTERY/OHM MEASUREMENT
1. Turn the function switch to AC, DC BATTERY, or OHM function

(7, 8, 9, or 11, Fig. 1).

2. Touch or connect leads to the desired circuit and read the display.


CONTINUITY/DIODE MEASUREMENT
1. Turn the function switch to the Continuity/Diode function (10 Fig. 1).
2. Press the Continuity Buzzer Sound On/Off button (6 Fig. 1), until the sound ON

signal is shown on the display “.)))”

3. For Continuity measurement connect the leads across the desired circuit and read

the display. If the impedance of the circuit is less than 20 ohms then the buzzer will
sound.

4. For Diode measurement connect the red test lead to the positive “+” pole of the

diode and the black test lead to the negative “-” pole of the diode. Read the forward
voltage drop on the display.

BATTERY READINGS
1. Turn the function switch to the Battery function (9, Fig. 1).
2. Touch or connect leads to the battery and read the display.
3. Incorporated in the circuitry is a resistive load to test batteries.

HOLD
Press the D-HOLD (Display-Hold) button (2, Fig. 1) to “freeze” the display and capture
a reading. Press it again to release the “freeze” and return to the current reading.

LOW BATTERY
The appearance of the low battery signal, “ ”, erratic readings, a faint display, or no
display are all indicators that the battery is low and needs to be replaced. To prevent
electrical shock, do not remove the battery when the test leads are in place. To
replace the battery, remove the battery cover located on the back of the unit and insert
two new 1.5V batteries with the positive side facing you (Control Company Cat. No.
1039) . Incorrectly installed batteries may damage electronics. Replace the Battery
Cover. To prolong the life of the unit, remove the batteries when the unit is not in use
for an extended period.

ALL OPERATIONAL DIFFICULTIES
If this multimeter does not function properly for any reason, please properly replace
the battery with two new 1.5V batteries (see Low Battery section, above). Low battery
power can occasionally cause any number of “apparent” operational difficulties.
Replacing the battery with a new fresh battery will solve most difficulties.

WARNINGS
Do not exceed 750 volts when in the AC or DC function. Do not exceed 500 volts
when in all other functions. OL on the display means over range. Before testing in
ohms or continuity/diode, OL will be on the display. To avoid electrical shock, never
touch your body to the metal part of the test lead when making a measurement. Do not
use the OHM or Continuity/Diode functions to test voltage. It may damage the meter’s
electronics.




WARRANTY, SERVICE, OR RECALIBRATION
For warranty, service, or recalibration, contact:

CONTROL COMPANY

4455 Rex Road

Friendswood, Texas 77546 USA

Ph. 281-482-1714 Fax 281-482-9448

E-mail [email protected]

www.control3.com

Control Company is ISO 9001 Quality-

Certified by DNV and ISO 17025 accredited

as a Calibration Laboratory by A2LA.

TRACEABLE

®

MULTIMETER INSTRUCTIONS