Control Company 4154 RH/TEMPERATURE MONITOR WITH ALARM User Manual
Control company

Traceable
®
is a registered trademark of Control Company ©2008 Control Company. 92-4154-00 Rev. 2 052308
 
SPECIFICATIONS 
Temperature 
Range: 
-58 to 158.0 °F / -50 to 70.0 °C
Resolution: 0.1° 
Humidity 
Range: 
20 to 99%
Resolution: 1% 
Battery: 
AAA (1.5V DC)
 
EXTERNAL TEMPERATURE AND HUMIDITY SENSOR 
The external sensor measures both temperature and humidity. Immersing or exposing 
the sensor to liquid will cause damage. 
 
 
 
 
DISPLAY MODES 
Temperature Display Mode: Indicated by the display of the current temperature and 
the MAX (maximum) and MIN (minimum) temperatures. 
Humidity Display Mode: Indicated by the display of the current humidity and the MAX 
and MIN humidity. 
To change from one mode to another press the MODE button. 
 
DISPLAYING °F OR °C 
To display the temperature readings in Fahrenheit or Celsius slide the switch on the 
back of the unit to °F or °C. 
 
RESETTING MIN/MAX MEMORIES 
 
Temperature MIN/MAX memories 
While in the Temperature Display Mode, press the CLR MEM button to clear the MAX 
and MIN temperature memories. The display will show “- -” under MAX and MIN and 
will then set the MAX and MIN to the current temperature reading. 
Humidity MIN/MAX memories 
While in the Humidity Display Mode, press the CLR MEM button to clear the MAX and 
MIN humidity memories. The display will show “- -” under MAX and MIN and will then 
set the MAX and MIN to the current humidity reading. 
 
ALARMS 
There are four alarm points that can be set: 
High Temperature Alarm-- 
Alarm will sound when temperature rises above this point 
Low Temperature Alarm-- 
Alarm will sound when temperature falls below this point 
High Humidity Alarm-- 
Alarm will sound when humidity rises above this point 
Low Humidity Alarm-- 
Alarm will sound when humidity falls below this point 
All four alarm set points are independent of each other. The user may enable/disable 
any or all of the alarm set points. 
 
Setting Temperature Alarms 
1. 
While in the Temperature Display Mode, press and hold the SET button until 
“HI” appears on the display 
2.
If “- - - -” appears on the display to the left side of “HI”, press the MODE button 
to enable the alarm. A temperature should now appear to the left of “HI” on the 
display. 
3.
To set the HIGH temperature alarm limit, press the CLR MEM/+ button to 
advance the display in 1° C increments until the desired high temperature is 
displayed. Press and hold the CLR MEM/+ button to rapidly advance the display. 
4.
Once the desired HIGH temperature is on the display, press the SET button. 
“LO” will now appear on the display. 
5.
If “- - - -” appears on the display to the left side of “LO”, press the MODE button 
to enable the alarm. A temperature should now appear to the left of “LO” on the 
display. 
6.
To set the LOW temperature alarm, limit press the CLR MEM/+ button to 
advance the display in 1° C increments until the desired low temperature is 
displayed. Press and hold the CLR MEM/+ button to rapidly advance the display. 
7.
Once the desired LOW temperature is on the display, press the SET button.
 
Setting Humidity Alarms 
1. 
While in the Humidity Display Mode, press and hold the SET button until “HI” 
appears on the display 
 
 
 
 
 
2. 
If “- - - -” appears on the display to the left side of “HI”, press the MODE button 
to enable the alarm. A number should now appear to the left of “HI” on the 
display. 
3.
To set the HIGH humidity alarm limit, press the CLR MEM/+ button to advance 
the display in 1% increments until the desired high humidity is displayed. Press 
and hold the CLR MEM/+ button to rapidly advance the display. 
4.
Once the desired HIGH humidity is on the display, press the SET button. “LO” 
will now appear on the display. 
5.
If “- - - -” appears on the display to the left side of “LO”, press the MODE button 
to enable the alarm. A reading should now appear to the left of “LO” on the 
display. 
6.
To set the LOW humidity alarm limit, press the CLR MEM/+ button to advance 
the display in 1% increments until the desired low humidity is displayed. Press 
and hold the CLR MEM/+ button to rapidly advance the display. 
7.
Once the desired LOW humidity is on the display, press the SET button.
 
Alarm Sounding 
The unit will sound the alarm in when the temperature or humidity rises above, or falls 
below the alarm set points; regardless of the display mode. The unit will continue to 
alarm as long as the current reading is above or below the alarm set point. Once the 
reading returns to an “in-range” condition, the alarm will stop sounding. To stop the 
alarm from sounding while above or below an alarm set point, the alarm set point must 
be disabled (see “Disabling an Alarm Set Point”). 
 
Disabling an Alarm Set Point 
To disable an alarm set point, follow the instructions for setting an alarm accordingly. 
When the desired “HI” or “LO” alarm set point is displayed, press the MODE button to 
disable the alarm. (Dashes “- - -” will appear to the left of “HI” or “LO” indicating that 
the alarm set point has been disabled.) The last High/Low alarm set point will be 
maintained in memory so that the set point can be enabled without having to be re-set. 
Each successive press of the MODE button will enable (number shown) or disable 
(dashes “- - -” shown) the alarm set point. 
 
BENCH STAND 
The unit is supplied with a bench stand that is a part of the battery cover. To use the 
bench stand, locate the small opening at the bottom of the unit. Place your fingernail 
into the opening and flip the stand up. To close the stand, simply snap it shut. 
 
WALL MOUNTING THE UNIT 
Set two screws into the wall at the desired location. Do not set the screw flush to the 
wall, the heads of the screws will need to slip into the receptacles on the back of the 
unit. Once the screws have been properly set, hang the unit in place by sliding the 
receptacles on the back of the unit over the heads of the screws. 
 
ALL OPERATIONAL DIFFICULTIES 
If this unit does not function properly for any reason, replace the battery with a new 
high-quality battery (see “Battery Replacement” section). Low battery power can 
occasionally cause any number of “apparent” operational difficulties. Replacing the 
battery with a new fresh battery will solve most difficulties. 
 
BATTERY REPLACEMENT 
Erratic readings, a faint display, or no display are all indications that the battery must 
be replaced. Slide the battery cover down in the direction of the arrow. Remove the 
exhausted battery and replace with a AAA alkaline battery. Make certain to insert the 
new battery with the proper polarity as indicated in the illustration in the battery 
compartment. Replace the battery cover. 
 
 
WARRANTY, SERVICE, OR RECALIBRATION 
For warranty, service, or recalibration, contact: 
CONTROL COMPANY
4455 Rex Road
Friendswood, Texas 77546 USA
Ph. 281-482-1714 Fax 281-482-9448
E-mail [email protected]
www.control3.com
Control Company is ISO 9001 Quality-
Certified by DNV and ISO 17025 accredited
as a Calibration Laboratory by A2LA.
TRACEABLE
®
RH/TEMPERATURE MONITOR WITH ALARM INSTRUCTIONS
--DO NOT IMMERSE THE EXTERNAL SENSOR INTO LIQUID--