Caron 6044 User Manual
Page 47
6024 & 6044 Series Operations Manual
Rev F 02/14/12
Page 47 of 56
SECTION 10
– SPECIFICATIONS
MODEL
6024
6026
6044
6046
Temperature
Range
10°C Above
Ambient to 60°C
5°C to 60°C
10°C Above
Ambient to 60°C
5°C to 60°C
Temperature
Control
± 0.1°C
Temperature
Uniformity
± 0.3°C
Temperature
Sensor
RTD
Humidity Range
Ambient to 95% RH
Humidity Control
± 3% RH
Humidity Sensor
Capacitive
CO2 Range
0-20% CO2
CO2 Control
± 0.1% CO2
CO2 Sensor
Infrared
Interior Dimensions
32" W x 27" D x 52.7" H
(81.3 cm x 68.6cm x 133.9cm)
32" W x 27" D x 66" H
(81cm x 69cm x 168cm)
Interior
Construction
Type 304, 2B Finish, Solid Stainless Steel
Exterior
Dimensions
35.5" W x 33.3" D* x 77.1" H
(90.2cm x 84.6cm x 195.8cm)
36" W x 33" D* x 90" H
(90cm x 85cm x 229cm)
Exterior
Construction
Cold Rolled Steel, Powder Coated
Work Space
25 Cu. Ft. (708 Liters)
33 Cu. Ft. (934 Liters)
# of Shelves
Four (4)
Five (5)
Shelf Construction
Type 304, Perforated Stainless Steel, Electropolished
Shelf Dimensions
29.25" W x 26.45" D
(74.3cm x 62.1cm)
Model
6024
6026
6044
6046
-1, -2
-3**
-1, -2
-3**
-1, -2
-3**
-1, -2
-3**
Shipping Weight
525 lbs
825 lbs
575 lbs
950 lbs
600 lbs
900 lbs
650 lbs
1025 lbs
Model
-1
-2
-3
Electrical
115V, 60Hz, 16A
230V, 60Hz, 10A
230V, 50Hz, 10A
Specifications are subject to change without notice.
Environmental Conditions: Temperature 15ºC to 25ºC, Humidity non-condensing
*Add 2.75 inches for handle
**Includes export shipping crate
*6020 Series units have forced internal air flow of 350 cfm (9,900 LPM)*