6024 & 6044 Series Operations Manual
Rev F 02/14/12
Page 22 of 56
Installing a Drain Water Pump (PUMP301)
In applications where a floor drain is not available and a CARON water recycling system
is not being used, a drain pump can be purchased to pump any excess condensate from
the incubator to a local sink or drain. The pump is located near the middle of the back of
the incubator. Connect the supplied tubing from the pump to the sink / drain. The
tubing may be run vertically into a ceiling but should not exceed 15 feet height.
Installing Side Mounted Recorders (RCDR305 & RCDR306)
The recorder will arrive packaged inside the chamber. Carefully remove the recorder
from its packaging. Mount the recorder by using the pre-installed recorder bracket.
There are three factory drilled holes located on the right side of chamber as you face the
front of chamber. Using the factory supplied screws, screw the recorder to the side of
the chamber.
There are two cables that come out of the recorder. One is to power the recorder; the
other is the temperature and/or humidity signals coming from the chamber. With power
to the equipment turned off, plug the two connectors into their mating connector at the
top of the chamber. Turn power to the chamber back on.
Standard factory set up for chart speed is 7 day operation. Refer to the Chart
Reco
rder’s User’s Manual provided with the recorder to change the chart speed settings
for various chart speeds.