beautypg.com

Caron 6044 User Manual

Page 32

background image

6024 & 6044 Series Operations Manual

Rev F 02/14/12

Page 32 of 56

The CO

2

gas supply should be 99.5% pure and

should not contain a siphon tube. Gas pressure
to the unit must be regulated to less than 30
psig. Failure to do so could cause tubing to
burst.

Built In Gas Guard System (GASG301 )

An optional built in gas guard system is available to allow two tanks of CO2 to be
connected to an incubator requiring approximately 15 psig of gas pressure. The unit is
designed to automatically switch from the primary tank to the secondary tank when low
gas pressure of approximately 10 psig is detected on the primary tank. This allows for a
continuous supply of CO2 to an incubator after the primary tank is empty. In addition,
the user is notified of a tank empty scenario via an audible and visual alarm.









The CO

2

gas supplies must be equipped with two stage regulators to ensure that the

incoming gas to the unit is regulated to appropriate levels. The high pressure stage
should have a 0-2000 psig range, and the low pressure gauge should adjust from 0-30
psig. When connecting the gas supplies, adjust each tank output to 20-25 psig. If the
appropriate regulators are not available, contact CARON customer service to purchase
them.

Once the cylinder regulators are installed and adjusted on each tank, connect the outlet
of the regulator on Tank 1 to the hose barb fitting labeled Tank 1 on the back of the unit.
Repeat the process for Tank 2. Turn on the regulated gas supplies and check the
connections closely for leaks.

The power switch is used to turn on/off the operation of the product. If power is off, the
unit will allow gas to flow through Tank 1 if required by the incubator.

The factory default “master tank” is Tank 1. When the appropriate gas pressure is
supplied to both tanks, the master tank will always be used as the gas source. The unit
will swap from the master tank to the alternative tank whenever a low gas pressure
condition is detected.

The alarm switch is used to disable the audible alarm when a low gas condition is
detected. The audible alarm can be silenced while a new tank is installed by selecting
the silence position. Once a low gas pressure condition is resolved, it is recommended
that the alarm switch be returned to audible to ensure audible notification of future low
gas pressure conditions. As a reminder, the Tank 1 and Tank 2 alarm lights will be
illuminated for as long as a low pressure condition is present on the respective tank.

This manual is related to the following products: