Caron 6044 User Manual
Page 23

6024 & 6044 Series Operations Manual
Rev F 02/14/12
Page 23 of 56
SECTION 5
–OPERATION
With the incubator properly installed and with water and CO
2
connected, the power
switch on the front lower left of the control bezel can be turned on. Within a few
minutes, the temperature, humidity, and CO
2
will begin moving toward their set-points.
Allow the unit to stabilize for 12 hours before use or prior to making any calibration
adjustments.
Changing the Temperature Set-point
If an alternative temperature set-point is required, the following steps can be taken:
Press the up arrow to increase the temperature set-point by 0.1
C. Press the down
arrow to decrease the temperature set-point by 0.1
C. Pressing and holding either
button will cause the set-point to scroll rapidly in either direction.
Temperature
Set-point
Actual
Temperature
Increase
Temperature
Set-point
Decrease
Temperature
Set-point
Decon Cycle
Active Indicator
Switch to
initiate Decon