6024 & 6044 Series Operations Manual

Rev F 02/14/12

Page 23 of 56

SECTION 5

–OPERATION


With the incubator properly installed and with water and CO

2

connected, the power

switch on the front lower left of the control bezel can be turned on. Within a few
minutes, the temperature, humidity, and CO

2

will begin moving toward their set-points.

Allow the unit to stabilize for 12 hours before use or prior to making any calibration
adjustments.


Changing the Temperature Set-point

If an alternative temperature set-point is required, the following steps can be taken:















Press the up arrow to increase the temperature set-point by 0.1

C. Press the down

arrow to decrease the temperature set-point by 0.1

C. Pressing and holding either

button will cause the set-point to scroll rapidly in either direction.

Temperature
Set-point

Actual
Temperature

Increase
Temperature
Set-point

Decrease
Temperature
Set-point

Decon Cycle
Active Indicator

Switch to
initiate Decon