Caron 6044 User Manual

Page 41

background image

6024 & 6044 Series Operations Manual

Rev F 02/14/12

Page 41 of 56

3. Lift the chart from the hub and locating pin and slide it from under the pens to remove
it from the chart plate.
4. Slip the new chart under the pen lifter, pens and time index; and press the chart into
place over the chart hub and locating pin.
5. Grasp the chart hub and locating pin and turn the chart until the desired time line on
the chart is aligned with the time index on the chart plate and Pen #1. Push down the
lifter to return the pens to the chart.
6. Close the door and apply power.




SHKR301/SHKR302/SHKR303 Multi Tier Shaker Support System

For applications that require shakers to be used in the incubator, a multi-tier reinforced
shaker support shelving system is available. This shelving allows the use of two or three
small shakers each to be used inside the incubator. Two-tier configurations have one
duplex outlet fused at 2A. Three-tier configurations have two duplex outlets fused at 4A
total.

Shakers inside incubators produce heat. To maintain control temperatures with 15C of
ambient, shakers that generated less that 50 watts each can be used. If high power
shakers are used in an incubator, a refrigerated unit such as CARON’s model 6026 or
6046 is required to remove the heat to allow the unit to control at lower temperatures.