6024 & 6044 Series Operations Manual

Rev F 02/14/12

Page 43 of 56

Calibrating the Humidity

If humidity calibration is needed, the following steps can be taken:















Locate the reference instrument’s temperature sensor in close proximity to the cabinet’s
geometric center. Be sure that the stabilization times described earlier have been
satisfied prior to performing this calibration. Press the advance key until the green
display reads CAL (calibrate). Pressing the up arrow will increase the Humidity
calibration offset by 1%. Pressing the down arrow will decrease the Humidity calibration
offset by 1%. Pressing and holding either button will rapidly scroll the calibration offset.
When finished, press the infinity key to return to the main menu.