Connect each incubator to a grounded circuit.
Failure to do so could result in electrical shock.
regulator to the hose barb fitting using the tubing and clamps provided. An inline HEPA
filter is provided to remove any contaminants in the CO2 gas supply. Check the
connections closely for leaks.
If the unit is equipped with a built in gas guard system, there will be 2 gas inlets. Each of
the inlets should be connected to an individual gas tank as described above.
Connecting Electrical Power
The unit requires a dedicated electrical outlet. See table below for model specific power
required and connection.
Model #
Power Requirements
Plug Connection
-1
115V, 60Hz, 16A FLA
NEMA 5-20
-2
230V, 60Hz, 10A FLA
NEMA 6-15
-3
230V, 50Hz, 8A FLA
CEE 7/7
When the required electrical connection is available, plug the provided power cord into
the unit and the electrical outlet.