beautypg.com

Caron 6044 User Manual

Page 15

background image

6024 & 6044 Series Operations Manual

Rev F 02/14/12

Page 15 of 56

Use only distilled or deionized water with a
resistivity between 50K

-CM and 1M

-CM and

a pH of greater than 6.5. Using water outside
this range will void your warranty.

Do not use water that contains chloramines.
Chloramines can damage internal rubber
gaskets resulting in leaks.

High concentrations of carbon dioxide can
cause asphyxiation. The use of CO

2

monitors

and alarms is recommended for areas where
CO

2

can collect.

The CO

2

gas supply should be 99.5% pure and

should not contain a siphon tube. Gas pressure
to the unit must be regulated to 15-20PSI.
Failure to do so could cause tubing to burst.

Connecting the Water Supply














To ensure proper operation, distilled or deionized water is required as a supply on units
that have humidity control. If these water sources are not available contact CARON
customer service.

A water inlet fitting

on the back of the unit and ¼” black tubing are provided to connect

the water supply to the chamber. Connect an appropriate water supply to the fitting.
Incoming line pressure should be regulated to not exceed 80 psi.

If a Condensate Recirculator water recycling system was purchased as a water supply,
refer to its user’s manual for proper installation of the water supply.

Connecting a CO

2

supply














The CO

2

supply should be 99.5% and not have siphon tubes. CO

2

pressure should be

regulated to 15-20 psi. CO

2

tank regulators can be purchased through CARON

customer service. Once the cylinder regulator is installed, connect the outlet of the

This manual is related to the following products: