Caron 7304 User Manual
Page 50
7300 Series Operations Manual
Rev A 3/26/2014
Page 50 of 80
If you press
the
(Close Window) button, the Alarm Window will close, but the
alarm will still be present as a flashing alarm icon on the status bar for the remainder of
the 60 minutes time. It will not reset the 60 minutes alarm countdown time if the alarm
condition is viewed on the pop up window.
After the 60 minutes time has passed for an alarm condition, the counter will reset itself
to 60 minutes and repeat the countdown process again until the alarm has been
resolved.
Muting the speaker:
By factory default, when an alar condition is present, the speaker will sound. This
speaker can be muted in an ‘on/off’ fashion eliminating all audible sounds. (Muting the
speaker will silence it until manually ‘un-muted’. This is different than ‘snooze’ in that
fact that snooze can only be enabled when an alarm condition is present and only lasts
for 1 hour.) When the speaker is muted, the alarm icon continues to flash and the
remote alarm contacts (optional) remain in the ‘alarm’ state.
To mute the speaker:
Press the
(Settings) button.
Settings
Button