beautypg.com

Caron 7304 User Manual

Page 19

background image

7300 Series Operations Manual

Rev A 3/26/2014

Page 19 of 80

Installing Drain Water Pump (PUMP301)

In applications where a floor drain is not available and a CARON water recycling

system is not being used, a drain pump can be purchased to pump any excess

condensate from the chamber to a local sink or drain. The pump is located near the

middle of the back of the chamber. Connect the supplied tubing from the pump to the

sink / drain. The tubing may be run vertically into a ceiling but should not exceed 15

feet height. The pump is equipped with a small reservoir on the bottom of the pump

with an internal level switch that will automatically turn the pump ON when it is full to

drain the water out of the reservoir and into a floor or sink drain.

Pump Outlet

to Sink or Floor

Drain

Pump Inlet

from Chamber

Drain

Reservoir with

Internal Level

Switch