Changing the Temperature Set-point
The steps below walk through an example of changing the temperature set point from
30.0 °C to 20.0 °C. This example shows humidity control as well (model 7000 series
only). Here is the display of the home screen.
To set the temperature set-point, press the
(Setpoint) button on the right
side of the screen.
On Plant Growth Chambers there is a Daytime and Nighttime Temperature setting
Once the Setpoint screen appears, press the
(Temperature Setpoint)
button. (In this example the temperature set point initially has a value of ’30.0’; this will
vary with different initial set point values.)
Actual
Temperature
Setpoint Button
Temperature
Setpoint Button