SECTION 7 – CALIBRATION
The temperature, lights (if dimming) and humidity (optional) systems can all be
calibrated as necessary. CARON recommends an annual calibration check of each
system. Before making a calibration adjustment, allow the cabinet to stabilize a
minimum of 12 hours from a power off condition. If the unit has been in operation,
allow a minimum of 3 hours of stable operation at all set-points. If you do not have the
appropriate reference instruments to perform calibration, contact CARON’s service
department for on-site calibration at
. Caron also provides
validation services which ensures that the unit is functioning properly according to IQ,
OQ and PQ protocols which satisfy FDA guidelines for qualification verification of
equipment.
The Calibration Screen
To get to the calibration screen from the home page:
Press the
(Settings) button.
Be sure that all reference instruments are
calibrated to an appropriate standard.
Settings
Button