7300 Series Operations Manual

Rev A 3/26/2014

Page 43 of 80

SECTION 7 – CALIBRATION

The temperature, lights (if dimming) and humidity (optional) systems can all be

calibrated as necessary. CARON recommends an annual calibration check of each

system. Before making a calibration adjustment, allow the cabinet to stabilize a

minimum of 12 hours from a power off condition. If the unit has been in operation,

allow a minimum of 3 hours of stable operation at all set-points. If you do not have the

appropriate reference instruments to perform calibration, contact CARON’s service

department for on-site calibration at

[email protected]

. Caron also provides

validation services which ensures that the unit is functioning properly according to IQ,

OQ and PQ protocols which satisfy FDA guidelines for qualification verification of

equipment.

The Calibration Screen

To get to the calibration screen from the home page:

Press the

(Settings) button.

Be sure that all reference instruments are

calibrated to an appropriate standard.

Settings

Button