7300 Series Operations Manual

Rev A 3/26/2014

Page 47 of 80

Calibrating Lights

If light calibration is needed, the following steps can be taken:

Be sure the reference instrument’s light sensor is a PAR (Photosynthetically Active

Raditation) sensor in units of µmole/m2/s. Place the reference instrument’s light

sensor on the same shelf and right next to the chamber’s light sensor approximately 12

inches (30 cm) beneath the lamps. Be sure that the stabilization times described

earlier have been satisfied prior to performing this calibration.

A positive value will move the light intensity ‘up’ and a negative value ‘down’. Press

the ‘home’ button and verify the proper light intensity is displayed. (See Temperature

Calibration Example.)

Calibrating Optional Chart Recorders

For calibrating the optional front and side mounted chart recorders, refer section 6

(Optional Accessory Operation)