7300 Series Operations Manual
Rev A 3/26/2014
Page 12 of 80
Leveling the Unit
Place a level on the middle shelf of the incubator. Adjust the feet until the unit sits level
left to right and front to back. Even if the unit is level without adjustment, the leveling
feet should still be lowered to avoid the cabinet moving while opening and closing the
outer door & prevent a flat spot from forming on the casters.
Connecting the Drain Line
The chamber drain connection is located in the bottom middle of the back of the
chamber. A 1/2” tube fitting and tubing are supplied in the unit parts kit. Slide the
tubing into the drain connection. Pull on the tubing after installation to make sure it is
tight. Route the drain tubing to a local floor drain.
If a local floor drain is not available, a variety of accessories are available through
CARON customer service. These accessories can also be viewed at
For non-humidified chambers operated above 15°C, a drain is not needed. Install the
1/2” plug into the drain connection.
When using a pressurized water source, failure to connect
the unit to a drain could result in facility flooding.
The drain line relies on gravity to remove water from the
chamber. The drain line must remain below the chamber
to drain properly. Kinks or elevations in the drain line
above the cabinet drain will not allow the chamber to drain.
Do not have multiple loaded shelves out
simultaneously or the chamber may tip.