beautypg.com

Caron 7304 User Manual

Page 18

background image

7300 Series Operations Manual

Rev A 3/26/2014

Page 18 of 80

Connecting Analog Outputs (OUTP301, OUTP302)

With the purchase of OUTP301 or OUTP302, the controls are equipped with analog

outputs. A set of terminals are provided on the back of the unit to connect to analog

outputs.

Analog outputs as either milliamps (4-20mA) or voltage (0-5V) signal output that

represents each of the displayed temperature (and humidity) values. These options

can be used for connection to in-house data acquisition, recorder, or alarm system.

Parameter

Voltage

Current

Corresponding Value

Temperature

0 – 5 V

4-20 mA

0 – 100 °C

Humidity

0 – 5 V

4-20 mA

0 – 100 %rh

CO2

0 – 5 V

4-20 mA

0 – 20 %CO2

Connect shielded wires to the appropriate signal terminals: I(+) for current (mA) or V(+)

for voltage (DC). For both current and voltage outputs, COM(-) is common terminal.

Analog Output

Connections