Replacement of the Pen:

Recorders that are equipped with fiber tipped cartridge pens will have a cartridge that

is color coded "red" to designate pen number one and an optional cartridge that is color

coded "blue" to designate pen number two. The pen cartridge is securely fastened

to the metal pen arm using a special "U" clip tab. For ease of replacement, it is

suggested that the two screws that hold the pen arm be loosened and the pen

cartridge and metal pen arm be removed as an assembly. Unsnap the plastic "U" clip

tab of the pen cartridge from the metal pen arm, remove and discard the old pen

cartridge. Replace the new cartridge by opening the hinge and snapping it securely

around the metal pen arm. Refer to the image below:

Pen Arm Calibration:

To check and/or adjust the recording pen(s) calibration to the outer most temperature

graduation of the chart, press and hold the "change chart" button (#3) until the pen

begins to move off of the chart. Once the pen(s) has moved off of the chart, again

press and hold the "change chart" button (#3) until the pen begins to move back onto

the chart. The pen should briefly stop at the outer most temperature graduation of the

chart before continuing onto the chart to begin recording. If the pen does not stop

exactly at this location on the chart, it can be adjusted or "calibrated" by using the left

(#1) or right (#2) arrow buttons.

When the pen moves back onto the chart and briefly stops, you will have

approximately five seconds in which to adjust the pen’s position using the left and right

arrow buttons of Figure 3.

On multiple pen recorders, each pen will move (one at-a-time) onto the chart briefly

stopping at the outer most temperature graduation of the chart at which time the pen’s

position can be adjusted by using the left (#1) or right (#2) arrow buttons. When the

time to adjust the position of the first pen has expired, the second pen will move onto

the chart briefly stopping at the outer most temperature graduation of the chart at

which time the second pen’s position may be adjusted.