beautypg.com

4 diaphragm vacuum pump for the complete system, 5 application parameters, Diaphragm vacuum pump for the complete system – Eppendorf Concentrator plus (New Design) User Manual

Page 41: Application parameters, Operating manual

background image

41

Concentrator plus/Vacufuge® plus — Operating manual

9.4

Diaphragm vacuum pump for the complete system

9.5

Application parameters

Max. current:

1.4/1.6 A (230 V, 50/60 Hz)
3.0 A (120 V, 50 Hz)
3.6/3.8 A (100 V, 50/60 Hz)

Max. power:

180 W (230/120/100 V)

Motor protection:

Thermal winding protection

Protection class according to IEC 529:

IP 54

Suction capacity (in accordance with DIN EN
60529; VDE 0470-1):

1.9/2.1 m

3

/h (50/60 Hz)

Achievable ultimate pressure (absolute):

<20 mbar

Max. permissible pressure at the outlet
(absolute):

2 bar

Max. differential pressure between inlet and
outlet:

1 bar

Nominal speed at 50/60 Hz:

1.500/1.800 rpm

Radio interference protection class

B

Materials of surfaces that are in contact with the medium

Inner part of housing cover:

PTFE, with carbon reinforcement

Head cover, diaphragm clamping disc:

ETFE, with carbon reinforcement

Valve:

FFKM

Diaphragm:

PTFE

Inlet:

ETFE

Outlet:

ETFE

Screw connection:

ETFE

Hose:

PTFE

Run time:

1 min to 9:59 h, infinitely (oo),
adjustable in 1 min increments.

Speed:

1,400 rpm, not adjustable

Max. relative centrifugal force
(rcf):

248 x g, not adjustable

Max. load:

96 micro test tubes of 2.0 mL each or
6 Falcon tubes of 50 mL each.

Permitted density of material for centrifuging
(at max. rcf/speed and max. load):

1.2 g/mL

Emission condenser volume:

280 mL

EN

O

perating manual