Oper at ing man u al – Eppendorf Concentrator plus (New Design) User Manual
Page 12

Concentrator plus/Vacufuge® plus — Operating manual
12
(1) Maximum payload per rotor bore for adapter + tube + content.
F-40-36-12
36 flat-bottom tubes of
1.5 mL each.
12 × 32
215 x g/
1,400 rpm
15 g
F-45-36-15
36 flat-bottom tubes of
3.0/5.0 mL each.
15 × 45-48
217 x g/
1,400 rpm
20 g
F-45-16-20
16 flat-bottom tubes
of 6.5/10.0 mL each.
20 × 42-55
217 x g/
1,400 rpm
45 g
F-40-18-19
18 flat-bottom tubes of
10.0 mL each.
19 × 66
228 x g/
1,400 rpm
35 g
F-45-12-31
12 flat-bottom tubes of
20.0 mL each.
31 × 55
226 x g/
1,400 rpm
55 g
F-35-8-24
8 flat-bottom tubes of
25.0 mL each.
24 × 86-90
232 x g/
1,400 rpm
90 g
A-2-VC
Two buckets to hold
•
0.2 mL PCR tubes
•
PCR strips of 8
and of 5
•
PCR plates
•
Microplates
•
Cell culture plates
•
Deepwell plates
(max. height =
27 mm)
128 Ч 86 Ч 27
131 x g/
1,400 rpm
115 g per bucket
•
Slides (with
CombiSlide
adapter)
26 × 75
Rotor
Tube
Max. capacity
Tube dimensions
(mm) Ш Ч L
and/or L × W × H
Max. g-force (rcf)/
speed (rpm)
Max. payload
per rotor bore
(1)
EN
Oper
at
ing man
u
al