Installation
4
Installation
4.1
Selecting the location
Select the location for the device according to the following criteria:
•
Suitable power connection as per the name plate (230 V/120 V/100 V).
•
Stable, horizontal and resonance-free lab bench. Weight of the device: 31/17 kg (complete
system/basic device).
•
A well ventilated environment which is protected from direct sunlight to prevent the device
from heating up more.
•
Ambient temperature during operation: 15 to 35°C.
•
At altitudes 1000 m above MSL (risk of insufficient cooling air supply), measures in
accordance with DIN EN 60034-1; VDE 0530-1 are required.
4.2
Preparing installation
Complete the following steps in the order described:
4.2.1
Open the packaging
1. Cut the adhesive tape.
2. Open up all four tabs of the box.
3. Remove the accessories and rotors.
4. Remove the cover cardboard.
5. Cut the strap retainers.
4.2.2
Lift the device out of the box
1. Lift the device and transport protection pad out of the box.
2. Place the device and transport protection pad on a stable, level and non-resonant lab bench.
NOTICE!
If a fault occurs, objects in the immediate vicinity of the device could get damaged.
In accordance with the recommendations of EN 61010-2-020, leave a safety distance of 30
cm clear around the device during operation.
NOTICE!
Damage from overheating.
Do not place the device close to sources of heat (e.g., radiator, drying cabinet).
Do not expose the device to direct sunlight.
Ensure free circulation of air by maintaining a distance of at least 30 cm on all sides of the
device from adjacent devices or the wall and keep the underside of the device clear.
Make sure that the air slots in the device are always free of obstruction.
Hint!
The weight of the device is 31.5 kg (complete system incl. gel dryer) or 16.5 kg (basic device).
•
Transport and installation of the device always requires 2 persons.
•
Only transport the device in the original packaging.
•
Use a transport aid for longer distances (e.g., hand cart).
•
Keep the packaging and transport securing devices for later transport or storage. See also
the instructions relating to transport (see p. 39).