2 error messages, 3 opening the device in case of a power failure, Error messages – Eppendorf Concentrator plus (New Design) User Manual
Page 38: Opening the device in case of a power failure, See p. 38), Oper at ing man u al
Concentrator plus/Vacufuge® plus — Operating manual
38
7.2
Error messages
If one of the following error messages appears, proceed as follows:
1. Remove fault (see Remedy).
2. Press the start/stop key to clear the error message.
3. If required, continue concentration and/or repeat.
7.3
Opening the device in case of a power failure
If the device lid cannot be opened during a power failure, you can activate the emergency lid
release manually.
1. Disconnect the power plug.
2. Insert a wire (max. 2.5 mm thick, e.g., a paper clip) into the opening on the right-hand side of
the housing (see point 5 in Fig. A and Fig. B on the fold-out page), and push against the
noticeable resistance.
This will release the device lid.
3. Open the device lid slightly.
4. Remove the wire.
5. Open the device lid fully.
Symptom / message
Cause
Remedy
Error 1
Drive fault
Contact Service.
Error 2
Error 3
Error 4
Drive blocked or stiff.
Move the rotor by hand, remove any
obstructions.
Check the max. load of rotor.
Check the rotor mounting.
Error 5
Drive fault
Contact Service.
Error 7
Error in the lid latch.
Contact Service.
Error 8
Fault at the temperature sensor
Contact Service.
Error 9
Electronics fault.
Caution! The rotor chamber
temperature may be > 72 °C.
Contact Service.
Error 10
Heater defective.
Contact Service.
Error 12
Error 14
Error 16
Electronics fault.
Contact Service.
IMBAL
Rotor unsymmetrically loaded.
Load the rotor symmetrically and balance.
WARNING!
Risk of injury from rotating rotor.
Wait for the rotor to stop before activating the emergency lid release.
Hint!
If there is a power failure, the vent valve will open. Before the rotor comes to a full stop, the
standard pressure in the rotor chamber will be restored.
EN
Oper
at
ing man
u
al