Concentrator plus/Vacufuge® plus — Operating manual
2.4
Rotors
You can operate the device with the following rotors. Before use of sample tubes, please note the
manufacturer's specifications with regard to centrifugation resistance (max. rcf).
Rotor
Tube
Max. capacity
Tube dimensions
(mm) Ш Ч L
and/or L × W × H
Max. g-force (rcf)/
speed (rpm)
Max. payload
per rotor bore
(1)
F-45-72-8
72 micro test tubes of
0.5 mL each.
8 × 31
inner ring:
202 x g,
outer ring:
224 x g/
1,400 rpm
10 g
F-45-70-11
70 micro test tubes of
1.5/2.0 mL each.
11 Ч 41/
11 Ч 47
inner ring:
153 x g,
middle ring:
202 x g,
outer ring:
248 x g/
1,400 rpm
15 g
F-45-48-11
48 micro test tubes
of 1.5/2.0 mL each.
11 Ч 41/
11 Ч 47
inner ring:
217 x g,
outer ring:
239 x g/
1,400 rpm
15 g
F-45-24-12
24 round-bottom tubes
of 6.0/8.0 mL each.
12 × 67-100
234 x g/
1,400 rpm
35 g
F-50-8-16
8 round-bottom tubes
of 15.0/20.0 mL each.
16 × 105-120
230 x g/
1,400 rpm
70 g
F-50-8-18
8 round-bottom tubes
of 15.0/20.0 mL each.
18 × 105-128
230 x g/
1,400 rpm
70 g
F-35-6-30
6 Falcon tubes
of 15 mL each.
17 × 116-123
232 x g/
1,400 rpm
25 g
F-45-8-17
8 Falcon tubes
of 15 mL each.
17 × 118-123
239 x g/
1,400 rpm
70 g
F-35-6-30
6 Falcon tubes
of 50 mL each.
29.5 × 116-123
232 x g/
1,400 rpm
75 g
EN
O
perating manual