Page 18: Safety instructions on the device, Oper at ing man u al

background image

Concentrator plus/Vacufuge® plus — Operating manual

18

3.5.5

Vacuum

3.6

Safety instructions on the device

WARNING!

Risk of injury from direct contact with the vacuum.

Do not expose any body parts to the vacuum of the device.

WARNING!

Risk of injury from overpressure.
Overpressure in the exhaust gas lines (e.g., from closed taps or blocked pipes) can cause the
pipes to burst.

Only use pipes with a sufficiently large cross section.

Always keep the exhaust gas line free.

Do not place any objects on the exhaust gas lines.

Do not bend the exhaust gas lines.

Do not fit any hose clamps or valves to the exhaust gas lines.

Note the maximum pressures and differential pressures permitted (see p. 41).

WARNING!

Risk to health from escaping substances.
No vapors from toxic liquids and pathogenic germs must escape.

Ensure that there is sufficient condensation and/or separation of the vapors via suitable cold
traps or chemical traps.

Please note the personal protective equipment (gloves, clothing, goggles, etc.), extraction
and safety class required for the laboratory.

Depiction

Meaning

Location

General hazard area

Follow the operating manual.

Complete system: side of the device next to the
mains connection.
Basic device: rear next to the mains
connection.

Danger of explosion

Do not use this device to process any
explosive, radioactive or highly reactive
substances.

Do not use this device to process any
substances which could create an explosive
atmosphere.

Top of the device

Risk of scalding when the device lid is open
When the heater is switched on, the surface
temperature of the rotor chamber can be > 60 °C.

Do not touch the wall of the rotor chamber.

Top of the device

Risk of injury from overpressure
The connection on the top of the emission
condenser is intended as an outlet only.

Make sure that the emission condenser is
connected correctly.

Never shut this connection off.

Top of the emission condenser

WARNING

5305 100.902-00

CAUTION

5305 100.910-00

CAUTION

Outlet for the
air flow on top.
Outlet only.

Do not cover.

5305 001.521-00