8 dispensing tool errors, 1 pipetting, Dispensing tool errors 13.8.1 – Eppendorf epMotion 5070 User Manual
Page 87: Pipetting
Technical data
epMotion
®
5070
English (EN)
87
13.8
Dispensing tool errors
13.8.1
Pipetting
Dispensing tool
Volume range
Pipetting
Testing
volume
Error limits
Error
Systematic error
Random error
± %
± μL
± %
± μL
TS 50
1.0 μL – 50 μL
1 μL
5 μL
25 μL
50 μL
±15.0
±5.0
±1.5
±1.0
±0.15
±0.25
±0.375
±0.5
±5.0
±3.0
±0.6
±0.4
±0.05
±0.15
±0.15
±0.2
TS 300
20 μL – 300 μL
20 μL
30 μL
150 μL
300 μL
±4.0
±3.0
±1.0
±0.6
±0.8
±0.9
±1.5
±1.8
±2.5
±1.5
±0.4
±0.3
±0.5
±0.45
±0.6
±0.9
TS 1000
40 μL – 1 000 μL
40 μL
100 μL
500 μL
1 000 μL
±5.0
±2.0
±1.0
±0.7
±2.0
±2.0
±5.0
±7.0
±1.5
±1.0
±0.2
±0.15
±0.6
±1.0
±1.0
±1.5
TM 50-8
1.0 μL – 50 μL
1 μL
5 μL
25 μL
50 μL
±25.0
±5.0
±2.0
±1.2
±0.25
±0.25
±0.5
±0.6
±10.0
±5.0
±1.2
±0.6
±0.1
±0.25
±0.3
±0.3
TM 300-8
20 μL – 300 μL
20 μL
30 μL
150 μL
300 μL
±10.0
±10.0
±2.5
±1.5
±2.0
±3.0
±3.75
±4.5
±4.0
±3.5
±0.8
±0.5
±0.8
±1.05
±1.2
±1.5
TM 1000-8
40 μL – 1 000 μL
40 μL
100 μL
500 μL
1 000 μL
±6.0
±3.0
±1.5
±0.8
±2.4
±3.0
±7.5
±8.0
±2.5
±1.5
±0.3
±0.15
±1.0
±1.5
±1.5
±1.5
- epMotion 96 (76 pages)
- epMotion 5075 (130 pages)
- Centrifuge 5427 R (64 pages)
- Centrifuge 5427 R (104 pages)
- White Paper 14 (8 pages)
- Rolling Cabinet (34 pages)
- Mastercycler nexus (118 pages)
- Mastercycler nexus (142 pages)
- Concentrator plus (New Design) (48 pages)
- Concentrator plus (43 pages)
- Easypet 3 (38 pages)
- Xplorer (74 pages)
- Xplorer Adjustment (26 pages)
- AF2200 Plate Reader (72 pages)
- AF2200 Plate Reader (78 pages)
- G0.5 µPlate (32 pages)
- BioSpectrometer basic (104 pages)
- BioSpectrometer kinetic (106 pages)
- BioSpectrometer fluorescence (102 pages)
- Micro Test Tubes (5 pages)
- Microplates (10 pages)
- PiezoXpert (34 pages)
- Eporator (38 pages)
- MiniSpin (20 pages)
- MiniSpin (25 pages)
- Centrifuge 5702 (32 pages)
- 5702 Centrifuge (27 pages)
- 5702 Centrifuge (32 pages)
- C5702 RH Centrifuge (32 pages)
- 5418 Centrifuge (80 pages)
- 5418 Centrifuge (48 pages)
- 5424 Centrifuge (71 pages)
- 5424 Centrifuge (44 pages)
- 5430 Centrifuge (88 pages)
- 5430 Centrifuge (130 pages)
- 5804 Centrifuge (95 pages)
- 5804 Centrifuge (127 pages)
- 5804 Centrifuge (129 pages)
- TransferMan4 r (102 pages)
- TransferMan4 m (96 pages)
- InjectMan 4 (100 pages)
- InjectMan NI 2 (60 pages)
- InjectMan NI 2 (16 pages)
- PatchMan NP 2 (53 pages)