Abb. 10-1: Removing the single-channel dispensing tool ejector
Fig. 10-1: Removing the single-channel dispensing tool ejector
8. Clean and decontaminate the dispensing tools.
The wipe decontamination of the dispensing tools is identical to the wipe decontamination of the device.
10.1.2
Carrying out spray decontamination
Prerequisites
• Decontamination agent in spray bottle
• Deionized water
1. Remove Labware and tools from the device.
2. Spray the surfaces with decontamination agent.
3. Allow the decontamination agent to take effect.
4. Wipe the decontamination agent with deionized water.
5. Allow the surfaces to dry.
NOTICE! Damage on the device due to ingress of liquids by spray decontamination.
Do not perform spray decontamination in and on the device. Use the decontamination
methods provided in the operating manual.