10 maintenance, 1 decontamination and cleaning, Decontamination and cleaning – Eppendorf epMotion 5070 User Manual
Page 71
Maintenance
epMotion
®
5070
English (EN)
71
10
Maintenance
10.1
Decontamination and cleaning
The device is cleaned as described in the Decontamination section.
Tab. 10-1: Decontamination methods and decontamination agents
WARNING! Danger due to flammable and infectious liquids in the waste container.
Pipette tips in the waste container may contain flammable or infectious liquids.
Wear personal protective equipment.
Handle the pipette tips and sample materials from the waste container in accordance with
the material safety data sheets, safety regulations and laboratory guidelines.
WARNING! Danger due to contamination of the LiquidWasteTub.
The outer surface of the LiquidWasteTub may become contaminated by pipette tips in the
waste container.
Wear personal protective equipment.
Decontaminate the outer surface of the LiquidWasteTub.
NOTICE! Material damage from corroding cleaning agents.
Cleaning and decontamination agents may contain corroding substances. Metal surfaces on
the epMotion become damaged by corrosion.
Use the cleaning agents and disinfectants specified in the operating manual.
Decontamination method
Decontamination agent
For use on
Wipe decontamination
70% (v/v) ethanol
Worktable
Waste container
LiquidWasteTub
Housing
Work surface adapter
Reflectors and sensor cover on the
Cleanbench
EasyCon Control-Panel
Labware
Tools
3%–4% sodium hypochlorite
solution
Waste container
LiquidWasteTub
Dispensing Tools
Racks
Thermoblocks
Thermoadapter
Height adapter
TipHolder