9.3.3.1 Stopping the application immediately

To stop the application immediately, e.g., at risk of collision, proceed as follows:

epMotion5070

 Open the front hood.

The application will be stopped immediately.

epMotion5070f

 Move the front screen of the Cleanbench up or down.

The front screen moves from the working position. The light barrier is interrupted. The application will
be stopped immediately.

epMotion 5070 and 5070f

 You can continue or abort the application.

9.3.3.2 Interrupting the application

 Stop the application using the

Stop application

button (refer to epMotion software operating manual).

The current movement will be ended. Then the application stops.

 You can continue or abort the application.

9.3.3.3 Continuing the application

Prerequisites

• The equipping of the worktable has not been changed.
• The levels of the vessels has not been changed.
• The carrier has not been moved by hand.

1. Close the safety screen.

2. Continue the application using the

Continue application

button (refer to epMotion software operating

manual).