Eppendorf epMotion 5070 User Manual
Page 68

Operation
epMotion
®
5070
English (EN)
68
9.3.3.1 Stopping the application immediately
To stop the application immediately, e.g., at risk of collision, proceed as follows:
epMotion5070
Open the front hood.
•
The application will be stopped immediately.
epMotion5070f
Move the front screen of the Cleanbench up or down.
The front screen moves from the working position. The light barrier is interrupted. The application will
be stopped immediately.
epMotion 5070 and 5070f
You can continue or abort the application.
9.3.3.2 Interrupting the application
Stop the application using the
Stop application
button (refer to epMotion software operating manual).
The current movement will be ended. Then the application stops.
You can continue or abort the application.
9.3.3.3 Continuing the application
Prerequisites
• The equipping of the worktable has not been changed.
• The levels of the vessels has not been changed.
• The carrier has not been moved by hand.
1. Close the safety screen.
2. Continue the application using the
Continue application
button (refer to epMotion software operating
manual).