4 completing your work, 1 cleaning up the worktable, Completing your work 9.4.1 – Eppendorf epMotion 5070 User Manual
Page 69: Cleaning up the worktable
Operation
epMotion
®
5070
English (EN)
69
9.3.3.4 Aborting the application
An application cannot be continued after it has been aborted.
Prerequisites
• The application was stopped or interrupted.
Abort the application using the
Abort application
button (refer to epMotion software operating manual).
•
The application will be aborted.
•
The tool holder returns the tool to its original location.
•
The carrier travels to the home position.
•
The epMotion is ready for operation.
You can start a new application.
9.4
Completing your work
9.4.1
Cleaning up the worktable
Complete the following steps after the end of the application.
1. Seal the vessels.
2. Remove the labware from the worktable.
3. Empty the waste container.
4. If necessary, decontaminate the device and labware (see Decontamination and cleaning on p. 71).
NOTICE! Contamination due to multiple use of pipette tips.
When you abort an application with multiple use of pipette tips, the epT.I.P.S. Motion
SafeRacks contain used pipette tips. The epMotion does not detect which pipette tips have
already been used.
Dispose of epT.I.P.S. Motion SafeRacks immediately after the application has been
aborted.
WARNING! Danger due to flammable and infectious liquids in the waste container.
Pipette tips in the waste container may contain flammable or infectious liquids.
Wear personal protective equipment.
Handle the pipette tips and sample materials from the waste container in accordance with
the material safety data sheets, safety regulations and laboratory guidelines.
WARNING! Danger due to contamination of the LiquidWasteTub.
The outer surface of the LiquidWasteTub may become contaminated by pipette tips in the
waste container.
Wear personal protective equipment.
Decontaminate the outer surface of the LiquidWasteTub.