1 carrying out wipe decontamination, Carrying out wipe decontamination – Eppendorf epMotion 5070 User Manual
Page 72
![background image](/manuals/551152/72/background.png)
Maintenance
epMotion
®
5070
English (EN)
72
10.1.1
Carrying out wipe decontamination
Prerequisites
• Decontamination agent
• Deionized water
• Lint-free cloths
• Cotton swabs
Cleaning the device
1. Switch off the device and disconnect it from the mains/power line.
2. Moisten the lint-free cloth and cotton swab with decontamination agent.
3. Clean the surface with a lint-free cloth. Clean hard to access surfaces with cotton swabs.
The surfaces are dampened with a decontamination agent.
4. Allow the decontamination agent to take effect.
5. Wipe the decontamination agent with deionized water.
6. Allow the surfaces to dry.
Cleaning the dispensing tools
7. Pull of the ejector from the single-channel dispensing tool.
Spray decontamination
outside the device
70% (v/v) ethanol
Waste box
LiquidWasteTub
Racks
Thermoblocks
Thermoadapter
Height adapter
TipHolder
UV radiation via Cleanbench
UV light of the Cleanbench
Irradiation period 15 min
If you are using UV irradiation
regularly, the epMotion must
be serviced annually by
Eppendorf Service.
Worktable without labware and tools
Autoclaving
121 °C
100 kPa overpressure
20 min
Waste box
LiquidWasteTub
Dispensing Tools
Racks
Thermoblocks
Thermoadapter
Height adapter
TipHolder
Decontamination method
Decontamination agent
For use on