IKA C 7000 Grundausstattung Set 2 User Manual
Page 50

3UHSDUDWLRQ DQG &DUU\LQJ 2XW 0HDVXUHPHQWV
IKA
-WERKE C 7000
Ver. 07 09.07
6HLWH
g
Here the user also has the choice of manual entry or automatic input from a con-
nected balance. Manual entries are made directly in this window using the alpha-
numeric keys, or automatic input is called up by pressing function key
I
. The mass
weighed by the balance is automatically accepted. After entry, confirm by pressing
the
↵
button. If you are preparing for a calibration, the LCD-display asks for the ref-
erence calorific value,
+VWDQG
If you are preparing for a calorific-value determina-
tion, the second line of the LCD-display briefly shows the enquiry
PHP VDYH
.
When it has disappeared, pressing the
PHP
button stores the data entered and
closes the entries for a calorific-value determination.
If you want to check the values before you store them, you can select each pa-
rameter using the arrow and function buttons and make corrections if necessary.
Once the parameters have been stored, it is no longer possible to change them.
If you are carrying out a calorific-value determination, omit the next step.
h
Only when preparing for a calibration: enquiry for reference calorific value,
+VWDQG
.
A charge-related pre-setting in the order of 26450 J/g will be displayed. This is the
calorific value of the benzoic acid tablets supplied by IKA
. If you are going to use a
different substance for calibration, use the key panel to enter the correct value; the
units must be J/g. Confirm by pressing the
↵
button. The second line of the LCD-
display briefly shows the enquiry
PHP VDYH
. When it has disappeared, pressing
the
PHP
button stores the data entered and closes the entries for a calibration.
The message "Bomb securely closed?" will appear. Ensure that the decomposition
vessel is properly closed and confirm with OK.
i
You can now place the decomposition vessel in the calorimeter. Use the carrying
and venting handle, C 7010.8. Place it on the decomposition vessel and turn it anti-
clockwise until the handle engages. Take care not to accidentally press the vent
knob, which will cause oxygen to escape. Guide the decomposition vessel into the
measuring cell, until it is standing vertically and centrally in the cell. Remove the
carrying and venting handle by turning it clockwise. Close the cover by drawing the
complete upper section of the instrument over the measuring cell. The LCD-display
then shows the text
VDPSOH FRGH
and the parameters just entered.
Activate
6WDUW
Each time 1000 ignitions have been performed using a given decomposition vessel,
the following message will appear:
1000 ignitions performed with Bomb x
Inspect decomposition vessel
or contact IKA service
This indicates that the decomposition vessel has reached a maintenance point and
that a safety check must be carried out. Confirm this message by pressing
7$%
and
2.
in sequence.
This message does not release the user from the responsibility of also continuously
checking the decomposition vessel for wear and carrying out safety inspections as
required.
☞