IKA C 7000 Grundausstattung Set 2 User Manual
Page 33

IKA
-WERKE C 7000
Ver. 07 09.07
3DJH
•
Sub-menu
67,QLW
UHDOW
The parameter
UHDOW
(real-time print out) is used to set the time interval in sec-
onds for sending measured values to a printer. The values 0, 3, 12, or 24 sec-
onds can be set. The preset value is 0 seconds . If the value 0 is selected, the
transmission of readings to a printer) during a test is switched off. The presenta-
tion of data on the LCD-display is unaffected by the setting of this parameter.
,I D SULQWHU LV QRW FRQQHFWHG WKH SDUDPHWHU UHDOW PXVW EH VHW WR
FDOFXO
In this sub-menu, you determine the calculation mode. Possible settings are:
VWG W
standard without titration
VWG W
standard with titration
FRDO W
carbon without titration
FRDO W
carbon with titration
The calculation mode must be set
EHIRUH
making a measurement, because sub-
sequent conversion to a different calculation mode is not possible.
The presetting is
VWDQGDUG ZLWKRXW WLWUDWLRQ
. The calculation modes
VWDQGDUG
and
FRDO
differ in that, when
FRDO
is set, the calculation of gross and net calorific val-
ues is carried out under various conditions. If
ZLWK WLWUDWLRQ
is selected, the en-
ergy of acid formation is determined by titration.
XQLW
The results for gross and net calorific values can be calculated and presented in
the output protocol in several different units. The corresponding settings for the
parameters
XQLW
are:
MRXOH
J/g
%78
BTU/lb
N:DWW
kWh/kg
FDO
cal/g
The preset unit is J/g.
SUH WLPH
In this menu, one of four values can be set for the pre-test time:
VHFRQGV
VHFRQGV
VHFRQGV
VHFRQGV
The preset value is
VHFRQGV
.
7KH ORQJHU WKH SUHWHVW WLPH WKH PRUH DFFXUDWH WKH UHVXOW
•
,*1 WLPH
The setting of
,*1 WLPH
determines the ignition time in milliseconds for the fixed
ignition wire. The works setting is 1000 ms, which is normally entirely adequate
to be sure that the cotton thread or the combustible crucible has ignited.
The preset ignition time should only be changed if the cotton thread or the com-
bustible crucible do not burn properly, or if platinum ignition wire is used. In
these cases, the ignition time should be increased in successive steps of not
more than 100 milliseconds. However, if combustion is incomplete, first check
the connections of the ignition wire to the electrodes in the decomposition ves-
sel.
☞
☞
☞