During transport and storage, the system must be protected against mechanical
shock, vibration, dust deposits, and corrosive atmospheres. In addition, the relative
humidity should not exceed 80%.
In case of repair the device has to be cleaned and free from any materials which
may constitute a health hazard.
If you require servicing, return the appliance in its original packaging. Storage
packaging is not sufficient. Please also use suitable transport packaging.
,QVWDOODWLRQ /RFDWLRQ
When installing the unit, observe the national and local regulations for the operation
of pressure vessels that apply at the site selected.
To ensure high accuracy of measurement, a constant ambient temperature is an
important precondition. For this reason, the chosen installation location should fulfil
the following conditions:
•
not exposed to direct sunlight;
•
no drafts (e.g. not beside a window, door or air-conditioning vent);
•
sufficiently far from heating radiators and other heat sources;
•
room temperature must be between 18 °C and 30 °C: to ensure good meas-
urement quality, there should be no variation in temperature;
•
the system must be installed on a horizontal surface;
•
If the C 7002 cooler is to be used, a water supply with a supply pressure of less
than 9 bar is required near the installation location, or a suitable cold water sup-
ply (e.g. IKA
KV 500).
For operation of the system, a power supply corresponding to the typeplates of
system components must be available at the installation location. An oxygen supply
with a pressure gauge, which can supply 99.95% pure oxygen, quality 3.5 at a pres-
sure of 30 bar is also required. A shut-off valve for the oxygen supply must be in-
stalled. Observe the instructions on handling oxygen given in Section 1 "For your
safety”.
☞