Notes

All of the digital displays have a built in lamp test feature. The password to
activate this function is:

L L R L L.

When activated, all segments on the display LED’s will illuminate for a few
seconds and then return to normal operation.

The digital displays that utilize a pressure sensor have a sensor check fea-
ture. The password to activate this function is:

L R R L R R.

Special Function Passwords:

Transducer Status Check:

L R R L R R

Pressure Gauge:

rP1.7 or above

Zero Calibration to match transducer to display:

L R R L L

Used if the gauge indicates pressure or vacuum erroneously.

Flowmeter:

Display Raw Pulses:

L R L R L L

Checks paddlewheel operation.

Noise cancellation:

L R L R R R R L

Cancels out flow indication from eddy currents on a transmitter when
the discharge valve is closed. Toggles

C on

or

C off.

Reset Totalizer:

L