Calibration for Breathing Oxygen Gauge installations:

Empty Cylinder (Closed Valve)

With the oxygen bottle closed and the system purged, enter
the calibration password.

L L L R R R

The left (0%) bar will flash to indicate that you are ready to calibrate for an
empty system.

Activate the left switch followed by the right switch.

The right (100%) bar will flash to indicate that the display is ready to calibrate
for a full system.

Maximum Volume

With a full oxygen bottle and the system pressurized, actuate the
right switch and then the left switch.

Calibration is complete.

NOTE:

Calibration should be accomplished using an oxygen
cylinder filled to the locally established maximum pressure.

Oxygen Supply