Problem

Possible cause Solution

The electrodes

have become

detached from

the skin.

The electrodes have lost their adhesive

power because they are dirty or

because they have been used for a

long time. You can try to clean the

electrodes (see chapter ‘Cleaning’).

If this does not help, replace the

electrodes. Please note that the

‘channel interrupted’ symbol only

disappears when you increase the

intensity.

The battery

symbol on

the menu

bar of the

remote

control

flashes.

This means

that the

batteries of

the remote

control are

low. The

battery status

indicator on

the TENS units

is solid orange.

You can continue treatment as long as

the treatment feels the same as usual.

When you press the on/off button

of the TENS units to stop treatment,

replace the batteries.

The

treatment

status

indicator on

the TENS

units flashes

red and the

treatment

has stopped.

The TENS

units

overheated

inside.

If the TENS units overheat inside,

remove them from the electrodes and

let them cool down. As soon as they

have cooled down sufficiently, you

can reattach them to the electrodes

and press the on/off button on the

TENS units to resume treatment. The

TENS units may become warm on

the outside during treatment, but the

temperature never becomes so high

that there is any risk of damage or

injury.