Problem
Possible cause Solution
The remote
control
does not
switch on.
The batteries
of the remote
control are
empty.
Replace the batteries (see chapter
‘Preparing for use’).
The batteries
are not
correctly
placed.
Open the battery compartment and
place the batteries correctly (see
chapter ‘Preparing for use’).
The remote
control may
be broken.
Contact the Philips Consumer Care
Centre.
The battery
status
indicator on
the TENS
units flashed
orange and
the device
switched off.
The battery of
the TENS units
is empty.
Charge the TENS units (see chapter
‘Preparing for use’).
The battery
status
indicator on
the TENS
units is solid
orange.
The battery of
the TENS units
is low. The
battery icon
on the remote
control flashes.
This means that the battery has 30%
of its capacity left. Recharge the TENS
units as soon as possible (see chapter
‘Preparing for use’).