Problem

Possible cause Solution

There is a

cross on

the display

and I cannot

increase the

intensity

level to

start a

treatment

(see chapter

‘Signals on

the device’).

The cross is

the ‘channel

interrupted’

symbol. It may

indicate that

the electrodes

are not placed

properly.

Make sure that the electrodes are

attached to the skin properly. Please

note that the ‘channel interrupted’

symbol only disappears when you

increase the intensity.

The ‘channel

interrupted’

symbol may

indicate that

the TENS unit

or units are

not placed on

the electrodes

properly.

Check if the TENS units are placed on

the electrodes correctly. Wireless TENS

Pro (PR3094) only: If the ‘channel

interrupted’ symbol is above the left

intensity button, check if the TENS units

are connected properly. If the cross is

above the right intensity button, check

if the connectors of the extension

cable are connected properly. Please

note that the ‘channel interrupted’

symbol only disappears when you

increase the intensity.

English

40