SECTION 8
TROUBLESHOOTING AND REPAIR
8.1 Before Considering Trouble
If the instrument is not operating properly for some reason, check it as follows:
•
Power is not turned on
Is the power cord loose ?
→
Plug in firmly.
↓
Is the fuse blown ?
→
Replace the fuse.
•
Synchronization is not established.
Are the transmit and receive interfaces the same ?
→
Check the set values and set them to the correct values.
(Termination conditions, output level, offset, etc.)
↓
Is the connection cable normal ?
→
Change the cable.
↓
Initialize the instrument. (Transmitter and receiver)
Set the receiver the same as the transmitter.
•
Error added
Is the cable loose ?
→
Retighten the connector.
↓
Is Error addition OFF ?
→
Set Error addition to OFF.
↓
Are the phase margin and bias margin sufficient?
→
Adjust so that the phase and offset are suitably cut.
•
The output waveform is distorted.
Is the cable loose ?
→
Retighten the connector.
↓
Is the CLOCK1 outputs terminated ?
→
use the attached cable to connect the outputs to the
↓
DUMMY terminal.
Are cables and connectors that have good high
→
Use cables and connectors that have good high fre-
frequency characteristics being used ?
quency characteristics.