3.5 Mounting (both CS4H and CS4L)
Warning
As the case is made of resin, do not use excessive force while screwing in the mounting bracket,
or the case could be damaged.
The torque is approximately 0.12 Nm.
To fulfill the Dust-proof/Drip-proof IP66 specification, mount this unit vertically and check the rigidity of the
panel where this unit is mounted. If rigidity is not enough, Dust-proof/Drip-proof IP66 specification may not
be guaranteed.
Mounting panel thickness: Within 1 ... 15 mm
Insert this unit from the front side of the panel. Attach the mounting brackets by the holes at the top and
bottom of the case and secure the controller in place with the screws.
(Fig. 3.5-1)
Up
Down
Up
Down