Dismounting, return and disposal, Faults causes measures – WIKA CPC2000 User Manual

Page 30

background image

30

WIKA operating instructions, model CPC2000

11155094.02 07/2012 GB/D/F/E

GB

9. Faults / 10. Dismounting, return and disposal

9. Faults

Faults

Causes

Measures

Instrument is not working,

the display is dark.

No voltage supply

Check whether the plug is

correctly in the socket

Battery is empty and must

first be recharged

Set pressure has not

been reached, the pump

is running continuously

System is not fully sealed;

to large hose diameter or

hose length

Slide hose on correctly;

eliminate any leaks

Hose diameter max. 6 mm

(see chapter 3 "Specifica-

tions“)

CAUTION!

If faults cannot be eliminated by means of the measures listed above, the

low-pressure controller must be shut down immediately, and it must be

ensured that pressure and/or signal are no longer present, and it must be

prevented from being inadvertently put back into service.

In this case, contact the manufacturer.

If a return is needed, please follow the instructions given in chapter

10.2 “Return”.

10. Dismounting, return and disposal

WARNING!

Residual media on the dismounted low-pressure controller can result in a

risk to persons, the environment and equipment. Take sufficient precau-

tionary measures.

10.1 Dismounting

Only disconnect test and calibration installations once the system has been depressu-

rised!

When disassembling the instrument the following must be followed:

1. Ensure that there is no positive or negative gauge pressure in the instrument and that

all parts of the instrument are at room temperature.

2. Switch the instrument off using the ON/OFF switch