WIKA operating instructions, model CPC2000

11155094.02 07/2012 GB/D/F/E

GB

6. Commissioning, operation

6. Commissioning, operation

Never perform functional tests with compressed or medical air. This can

damage instruments with low pressure ranges.

Protect the instrument from sunlight, as this can lead to measuring

errors.

6.1 Operating mode 'Control' (CTRL)

The CTRL mode is used for the calibration of pressure sensors and other pressure

instruments. In CTRL mode, the pump is active and regulates according to the desired

pressure. The internal sensor provides the current value on the display.

Setting

Press the MENU button. Confirm the menu point MODE with SELECT and press the

/

button until CTRL appears. Confirm with SELECT.

In the submenu RANGE the upper limits of the pressure range are defined

(

/

button and / buttons) and confirmed with SELECT

In the submenu UNIT the pressure units are defined (

/

buttons) and confirmed

with SELECT .

In the submenu STEP the step sequence in % is defined (

/

buttons and /

buttons), confirmed with SELECT.

With the MENU button, you can exit the submenu

Operation

The instrument will revert to VENT mode

on exiting the menus, meaning the

sensor of the calibrator will be vented. In

the top centre of the display, the set point

of the pressure range will be displayed.

In the bottom centre of the display, the

percentage value can be read. On exiting

the menus, this value will be 0 %. With

the SELECT button, one can switch

between the set pressure value and the

percentage value.

The percentage value can be altered with the

/

buttons (in the defined step

sizes) and altered with the / buttons.

Change set point by x %

Pressure change in % via

/

buttons

Change pressure to 100 % via button

Change pressure back to 0 % via button