WIKA operating instructions, model CPC2000

11155094.02 07/2012 GB/D/F/E

GB

4. Design and function

The front screen of the display, behind the membrane, is made from

glass. If there is any possibility of the window breaking during operation

(including damage to the front membrane), all personnel in the vicinity of

the instrument must wear eye protection before and during its use.

4.5.1 On/Off button

Button to turn the instrument on and off.

4.5.2 Menu button

With this button the SETUP menu can be accessed, and there the

settings for each operating mode can be made. When the CTRL mode

is used, you will find that the device is left in VENT mode on leaving the

SETUP menu.

The SETUP menu items are made up of six submenu items. For a detailed description of

these items, see chapter 6.5 „SETUP settings (SETUP)“.

4.5.3 LEAK button

With the LEAK button, a device connected to the controller can be

tested for leaks. (This feature is only available in CONTROL mode.)

4.5.4 ZERO button

Button for manually correcting the zero point of the integrated sensor.

External influences such as temperature, pressure, mounting position or

surroundings, can alter the zero point of the instrument, i.e. the display

when the pressure ports are open. During zero adjustment, the device

automatically detects these changes and corrects them in the current

pressure display. The zero point adjustment requires a change of the

internal valves, which is not possible without loss of pressure.

(If this should interfere with the application, the automatic zero point

adjustment can be switched off.)