Gb 7. interface, Command answer function – WIKA CPC2000 User Manual

Page 27

background image

27

WIKA operating instructions, model CPC2000

11155094.02 07/2012 GB/D/F/E

GB

7. Interface

Command

Answer function

Control mode
:pa <-110 ... 110>

Increase the set pressure demand by x percent

-110 ... +110 --> change in set pressure demand in %

:pd

Decrease set pressure demand by currently-set step size in %

(Step DOWN).

:pr <-1100 ... 11000>

Adjust the current working and measuring ranges

-1,100 ... +11,000 --> new measuring range in 0.01 % FS

:ps <-10 ... 110>

Percentage set pressure demand -10 ... +110 --> set pressure

demand in %

:pu

Increase set pressure demand by currently-set step size in %

(Step UP).

:saz<0 | 1>

Set auto zero (in measuring and control modes)

0 --> off

1 --> on

:sbr<0 ... 1>

Set baud rate RS-232

0 --> 1200

1 --> 2400

2 --> 4800

3 --> 9600

4 --> 14400

5 --> 19200

6 --> 28800

7 --> 38400

8 --> 56000

9 --> 57600

:sbu<0 ... 1>

Set baud rate USB

0 --> 1200

1 --> 2400

2 --> 4800

3 --> 9600

4 --> 14400

5 --> 19200

6 --> 28800

7 --> 38400

8 --> 56000

9 --> 57600

:sci

Select active interface

Set communication interface

n --> Interface off

u --> USB active

r --> RS-232 active

sdb <0 ... 100>

Display brightness

0 ... 100 --> brightness in %

:spu <0 ... 9>

Pressure unit

0 --> Pa

1 --> hPa

2 --> ka

3 --> mbar

4 --> bar

5 --> Torr

6 --> mmHg

7 --> inHg

8 --> psi

9 --> mmH

2

O

10 --> inH

2

O

:ssl

System language

d = German

e = English