beautypg.com

WIKA LH-20 User Manual

Page 25

background image

25

8 Supplement

WIKA Operating instructions - Submersible pressure transmitter model LH-20

Voltage supply
Operating voltage 4 … 20 mA

Ʋ Deviation < 0.2 %

8 … 30 V DC

Ʋ Deviation < 0.1 %

9.6 … 30 V DC

Operating voltage 4 … 20 mA/HART

9.6 … 30 V DC

Permissible residual ripple

Ʋ < 100 Hz

U

ss

< 1 V

Ʋ 100 Hz … 10 kHz

U

ss

< 10 mV

Reverse voltage protection

Integrated

load max.

≤ (U

B

-10 V)/0.022 A

Integrated overvoltage protection
Nominal leakage current (8/20 µs)

5 kA

Min. response time

< 25 ns

Electrical protective measures
Protection transmitter

IP 68 (30 bar)

Overvoltage category

III

Protection class

III

CE conformity
ATEX 94/9/EG (optional)
4 … 20 mA:
EMC directive 2004/108/EG, EN 61326 Emission (group 1, class B) and interference resistance

(industrial area)
4 … 20 mA/HART:
EMC directive 2004/108/EG, EN 61326 Emission (group 1, class A) and interference resistance

(industrial area)

Warning:

according to EN 55011 or CISPR 11
This is an equipment of class A for emitted interference and is designed for use in

industrial environment. In other environments such as e.g. in living or business areas, this

equipment can interfere other instruments.