change on water balance.

The LSI titration function generates an accurate saturation index value

that indicates the state of the system at the time the calculation is

made.

C. Hardness Units

nOte: The 9P MUST be in the HARDNESS FUNCTION or the

hardness screen (“hrd”) in LSI CALCULATOR or LSI TITRATION

mode for hardness unit preference selection.

The hardness unit preference is set simultaneously for all three

functions that capture a hardness value. That means when you set the

preference in one function, it is set for the other two functions at the

same time.

The LSI Calculator allows you to select either ppm or grains of

hardness units.

The hardness unit conversion is based on the following equivalency:

17.1 ppm (mg/L) = 1 grain

XXv. sOFtware versiOn

Contact the Myron L Company to see if a software upgrade is available.

1.

Press any parameter key.

2.

Press

key until three numbers are displayed as shown

in Figure 35.

3.

Press any parameter key; instrument will time out in ~15

seconds.