25

pg. 29) is performed separately for each type of solution one wishes to

measure (ref. Conductivity/TDS Standard Solutions, pg. 56).

D. Procedure to Select a Solution

nOte: Check display to see if solution displayed (KCl, NaCl, 442 or

User) is already the type desired. If not:

1.

Press

,

or

to select the parameter on

which you wish to change the solution type.

2.

Press and hold

key for 3 seconds to make “sel

appear (see Figure 1). For demonstration purposes, all 4

solution types are shown simultaneously.

3.

Use the

or

key to select type of solution desired

(ref. Solution Characteristics, pg. 61). The selected solution

type will be displayed: KCl, NaCl, 442 or User.

4.

Press

to accept new solution type.

E. Application of User Solution Type

1. User Programmable Temperature Compensation

(Tempco)

This feature allows you to change your Ultrameter III’s temperature

compensation factor to another factor between 0-9.99%/°C (ref.

Temperature Compensation, pg. 58). This feature does not apply to pH

or ORP.

a.

As in Procedure to Select a Solution, above, select User mode.

b.

With User mode now selected, press

. You may now

adjust a temperature compensation from .00%/°C to 9.99%/°C,

by pressing

or

. See example in Figure 2, pg. 26.